Details
Specifications The TR125 Reactor is for chemical digestion of samples that require it, such as in the determination of COD, TOC, Total Phosphate and Total Nitrogen.The required temperatures and reaction times are pre-programmed, and easy to select by using the keypad. Users can pre-select reaction times of 30, 60 and 120 minutes or can set it to operate continuously. When digestion is complete, the reactor automatically turns off and will sound an alarm.Instrument supplied with:ReactorPlastic ShieldPower Adapters: US, EU, UK stylesOperating instructions & guarantee certificateHousingRound metal block in epoxy covered housing, 24 holes 16 mm diameterKeypadRobust membrane keypad with 4 keysTemperature Settings100 / 120 / 150 °C (± 0.3 °C)Timer30, 60, 120 minutes or continuous operation. Automatic turn-off with alarmHeating400 Watts, electronically controlled, with overheating protectionPrep. Timeapprox. 10 minutesPower Supply230 V /115 V /50-60 HzWeight7.9 poundsDimensions10.83 x 6.10 x 3.74 inches (LxWxH)CertificationCE